ABSTRACT
Introduction. The Diagnostic Reference Levels (DRLs) is defined as the reference values for typical radiation doses in specific imaging procedures. This study focuses on establishing Diagnostic Reference Levels (DRLs) for head CT scans at the Garoua Regional Medical Imaging Center. Methodology. The data were collected at the regional medical imaging center of the north. Radiation dose data were collected from the local ct scanners ge revolution ct 65-slice, ge healthineers, europe -central and eastern) over a period of 12 months from january to december 2023. And analyzed using microsoft excel 2016 and spss software version 23. The study explores imaging protocol parameters, such as voltage, current, pitch, and scan length, providing comprehensive insights into the technical aspects of head CT procedures. Results. Over a 12-month period, data from 621 patients were analyzed, revealing a male predominance with 52.5% and the most represented age group was 40 to 59 years old. the stroke protocol was the most used one with trauma as indication with more than 27.2% of the Head CT examination. For stroke and trauma, a voltage of 120 Kv, variable current (265 mAs for trauma), pitch range (0.52) Conclusion. DRLs are not the suggested or ideal dose for a particular procedure, but rather represent the level at which an investigation of the appropriateness of the dose should be conducted. They are supplements to professional judgment and are intended to enable individual CT users and the community at large to identify and address consistently high doses.
RÉSUMÉ
Introduction. Les Niveaux de Référence Diagnostiques (NRD) sont définis comme les valeurs de référence pour les doses de radiation typiques dans des procédures d'imagerie spécifiques. Cette étude se concentre sur l'établissement des Niveaux de Référence Diagnostiques (NRD) pour les scanners cérébraux par CT au Centre Régional d'Imagerie Médicale de Garoua. Méthodologie. Les données ont été recueillies au centre régional d'imagerie médicale du nord. Les données de dose de radiation ont été collectées auprès des scanners CT locaux (ge revolution ct 65-slice, ge healthineers, europe - central and eastern) sur une période de 12 mois, de janvier à décembre 2023. Et analysées à l'aide de Microsoft Excel 2016 et du logiciel SPSS version 23. L'étude explore les paramètres de protocole d'imagerie, tels que la tension, le courant, le pas et la longueur de balayage, fournissant des informations complètes sur les aspects techniques des procédures de CT cérébrales. Résultats. Sur une période de 12 mois, les données de 621 patients ont été analysées, révélant une prédominance masculine avec 52,5% et le groupe d'âge le plus représenté était celui des 40 à 59 ans. Le protocole d'accident vasculaire cérébral était le plus utilisé, avec un traumatisme comme indication dans plus de 27,2% des examens de CT cérébrale. Pour l'accident vasculaire cérébral et le traumatisme, une tension de 120 Kv, un courant variable (265 mAs pour le traumatisme), une gamme de pas (0.52). Conclusion. Les NRD ne sont pas la dose suggérée ou idéale pour une procédure particulière, mais représentent plutôt le niveau auquel une investigation de l'adéquation de la dose devrait être effectuée. Ils sont des compléments au jugement professionnel et ont pour but de permettre aux utilisateurs individuels de CT et à la communauté en général d'identifier et de régler les doses élevées de manière cohérente.

Keywords

Niveau Reference Diagnostique imagerie médicale Diagnostic Reference level Head Ct Scan

Article Details

How to Cite
Mohamadou Aminou, Onana Yaninck R, Boutche Laurent Y, Mbozo’o Mvondo S, Kouam Foubi BB, Djamilatou Oumarou, Balkissou Adamou D, & Mbo Amvene Jeremie. (2024). Diagnostic Reference Levels of Head CT Scan at Garoua Regional Medical Imaging Center. HEALTH SCIENCES AND DISEASE, 25(2). https://doi.org/10.5281/hsd.v25i2.5207

References

  1. - Alrehily F. Diagnostic reference levels of radiographic and ct examinations in Saudi arabia: a systematic review. Radiat Prot Dosimetry. 2022 Oct 16;198(19):1451-1461. doi: 10.1093/rpd/ncac183. PMID: 36125219.
  2. - Korir, G.K., Wambani, J.S., Korir, I.K., Tries, M.A. and Boen, P.K. (2015) National
  3. - B. Moifo et al. Diagnostic Reference Level Initiative for Computed Tomography Examinations in Kenya. Radiation Protection Dosimetry, 168, 242-252.
  4. - Moifo, B., Moulion Tapouh, J.R., Neossi Guena, M., Ndzana Ndah, T., Samba, R.N. and Simo, A. (2017) Diagnostic Reference Levels of Adults CT-Scan Imaging in Cameroon: A Pilot Study of Four Commonest CT-Protocols in Five Radiology Departments. Open Journal of Medical Imaging, 7, 1-8. https://doi.org/10.4236/ojmi.2017.71001
  5. - Cordoliani, Y.-S. (2010) Les enjeux de la radioprotection en imagerie médicale: The Issues of Radiation Protection in Medical Imaging. Journal de Radiologie, 91, 1184- 1185. https://doi.org/10.1016/S0221-0363(10)70170-0
  6. - Vano E, Miller DL, Martin CJ, Rehani MM, Kang K, Rosenstein M, et al. ICRP publication 135: diagnostic reference levels in medical imaging. Ann ICRP 2017;46(Oct):1e44
  7. - Treier R, Aroua A, Verdun FR, Samara E, Stuessi A, Trueb PR. Patient doses in CT examinations in Switzerland: implementation of national diagnostic reference levels. Radiat Protect Dosim 2010; 142(2e4):244e54. https://doi.org/10.1093/ rpd/ncq279
  8. - UK Health Security Agency. National Diagnostic Reference Levels (NDRLs) from 19 August 2019. Available from: https://www.gov.uk/government/publications/
  9. - Dawd JE, Ozsahin DU, Ozsahin I. A Review of Diagnostic Reference Levels in Computed Tomography. Curr Med Imaging. 2022;18(6):623-632. doi: 10.2174/1573405617666210913093839. PMID: 34517807.
  10. - Roberto M Sanchez et al 2022 J. Radiol. Prot. 42 041002DOI 10.1088/1361-6498/aca2b3
  11. - Sanchez RM, Siiskonen T, Vano E. Current status of diagnostic reference levels in interventional cardiology. J Radiol Prot. 2022 Dec 14;42(4). doi: 10.1088/1361-6498/aca2b3. PMID: 36379055.
  12. - Zamani H, Kavousi N, Masjedi H, Omidi R, Rahbar S, Perota G, Razavi E, Zare MH, Abedi-Firouzjah r. estimation of diagnostic reference levels and achievable doses for pediatric patients in common computed tomography examinations: a multi-center study. Radiat Prot Dosimetry. 2021 Aug 2;194(4):214-222. doi: 10.1093/rpd/ncab093. PMID: 34244802.
  13. - Diagnostic reference levels in medical imaging: review and additional advice. Ann ICRP. 2001;31(4):33-52. PMID: 12685758.
  14. - Thomas P. National diagnostic reference levels: What they are, why we need them and what's next. J Med Imaging Radiat Oncol. 2022 Mar;66(2):208-214. doi: 10.1111/1754-9485.13375. PMID: 35243776.
  15. - Tabesh J, Mahdavi M, Haddadi G, Ravanfar Haghighi R, Jalli R. Determination of Diagnostic Reference Level (DRL) in Common Computed Tomography Examinations with the Modified Quality Control-Based Dose Survey Method in Four University Centers: A Comparison of Methods. J Biomed Phys Eng. 2021 Aug 1; 11(4):447-458. doi: 10.31661/jbpe.v0i0.2105-1322. PMID: 34458192; PMCID: PMC8385214.

Most read articles by the same author(s)